| MIL-C-14858A (AR)
AMENDMENT 3
APPENDIX A
ANNEX 2
e. In all installations, the transducer shall be tightened with a torque wrench to 180 in-
lbs.
f. The signal line shall be reconnected to the transducer and securely tightened, finger
tight.
g. For calibration of the Kistler 6215 transducer, the diaphragm protection Type 6567
and second sealing ring Type 1100 shall not be used.
h. Transducer storage Transducers should be stored in a clean, moisture-free
environment. Transducers, diaphragm protections and cables shall be kept free from
contamination and possible damage when not in use.