| MIL-R-85523A(AS)
TABLE III.
Steady state outlet pressure conditions and requirements
(see 3.6.2.2.1).
Outlet
Altitude
Outlet
flowrate
pressure
(liters/min)
(x1000 ft)
(inches H2O)
0, 50
0,
15,
30
0.5
to
1.5
75
0,
15,
30
0.7
to
2.0
100
0,
15,
30
1.0
to
2.7
180
0,
15,
30
2.0
to
4.5
0, 50
34
0.5
to
2.7
75
34
0.7
to
3.2
100
34
1.0
to
3.7
180
34
2.0
to
5.7
0,
50,
75, 100,
180
36
3.0
to
6.0
0,
50,
75, 100,
180
40
8.0
to
11.0
0,
50,
75, 100,
180
45
13.0
to
16.0
0,
50,
75, 100,
180
50
16.0
to
20.0
Note:
1. The regulator shall be tested at inlet pressures of 5,8,15,30,90 and 120
psig at each of the above conditions, except as follows:
a.
At an inlet pressure of 5 psig, flowrates above 75 lpm shall not be
required.
b.
At an inlet pressure of 8 psig, flowrates above 100 lpm shall not be
required.
2.
Fluctuations in output flow shall not be permitted.
3. The conditions and requirements listed above pertain to first article and
qualification testing as described in 4.8.2.2.1 and 3.6.2.2.1, respectively.
Conditions and requirements for quality conformance inspection, quality
conformance verification inspection, environmental exposure testing, and post
environmental exposure are listed in Table IV.