TABLE IV.
Steady state outlet pressure conditions and requirements
(see 3.6.2.2.2).
Outlet
Altitude
Outlet
flowrate
pressure
(liters/min)
(x1000 ft)
(inches H2O)
0
0, 30
0.5
to
1.5
75
0, 30
0.7
to
2.0
180
0, 30
2.0
to
4.5
0
34
0.5
to
2.7
75
34
0.7
to
3.2
180
34
2.0
to
5.7
0, 75, 180
36
3.0
to
6.0
0, 75, 180
45
13.0
to
16.0
0, 75, 180
50
16.0
to
20.0
Note:
1. The regulator shall be tested at inlet pressures of 8,30 and 120 psig at
each of the above conditions, except as follows:
a.
At an inlet pressure of 5 psig, flowrates above 75 lpm shall not be
required.
b.
At an inlet pressure of 8 psig, flowrates above 100 lpm shall not be
required.
2. The conditions and requirements listed above pertain to the following
testing.
a.
Quality conformance inspection.
b.
Quality conformance verification inspection.
c.
Environmental exposure inspection.
d.
Post-environmental inspection.
3. First article and qualification test conditions and requirements of
4.8.2.2.1 and 3.6.2.2.1 are listed in Table III.
4.
Fluctuations in output flow shall not be permitted