MIL-T-62299B(AT)
The laboratory area shall include the
3.5.9.4 Laboratory area.
following:
(a) A Formica, or equal, surfaced work counter shall be installed
above the generator compartment at the rear of the vehicle.
(b) One blood plasma freezer, Puffer-Hubbard Model BVF-7 or equal,
shall be provided and secured to the rear work counter.
(c) One blood bank refrigerator, Puffer-Hubbard Model BB-7 or
equal, shall be provided and secured to the rear work counter.
(d) Hinged Formica, or equal, surfaced work counters shall be
installed adjacent to the rear work counters on the streetside
and the curbside.  The streetside counter shall be securely
mounted to the wall, shall hinge down from the wall to a
working level 36 inches from the floor, and shall measure 36
inches in length and 30 inches in depth.  The curbside counter
shall be securely mounted to the rear access door (see (h)),
shall hinge down to a working level 36 inches from the floor,
and shall measure 30 inches in length and 30 inches in depth.
The counters shall be held in up position by positive holding
devices and shall be adequately supported when in working
position.
(e) One DuPont Sorvall Model RC-3B with rotor, or equal, automatic
refrigerated centrifuge shall be installed next to counter on
the streetside.  The centrifuge shall be provided with one
extra set of cups.  Centrifuge unit shall include elevated
control panel with combined master switch and circuit breaker,
speed selector permitting automatic acceleration to all speeds
from 150 rpm to 6,000 rpm, gyro-action self-centering direct
drive , heavy duty refrigeration system with temperature
control range from minus 20 degrees Celsius ("C) to plus 40C
at full speed, adjustable dynamic braking, electric
tachometer, pilot lights, temperature control and indicator.
Units to operate from 230 volts ac, 60 Hz, single phase, 25
ampere vehicle supply.  Units shall be equipped with 30-minute
timer.
(f) Walls within room shall be surfaced with nylon, industrial
grade carpet with foam back to increase sound-deadening
qualities.  The carpet furnished shall be the same as
specified in 3.5.4.2.
(g) Six 120 volt exhaust fans shall be installed within the walls
for heat dissipation.  Two fans shall be located within the
curbside wall, one behind the centrifuge and one below the
hinged work counter on the streetside wall, and one each
behind the refrigerator and the freezer within the rear wall.
(h) Double access doors, providing an opening of not less than 60
inches by 60 inches, shall be installed on the curbside for
removal of equipment.