4.8.5.2.1 Cold weather operations (high). A fresh battery shall be
installed in the ventilator and the battery compartment shall be closed. The
ventilator shall be placed in the climatic chamber maintained at 122 degrees F
(50 degrees C). The ventilator motor shall be switched "ON" and operate for 4
hours . The minimum flow measured at the end of the delivery hose shall be as
specified in 3.10.4.3.2.
4,8.5.2.2 Cold weather operations (1ow). A fresh battery shall be
installed in the ventilator and the battery compartment shall be closed. The
ventilator shall be placed in the climatic chamber maintained at -4 degrees F
(-20 degrees C). The ventilator motor shall be switched "ON" and operate for
4 hours. The minimum flow measured at the end of the delivery hose shall be
as specified in 3.10.4.3.2.
5 PACKAGING
Preservation shall be level A or Commercial (see 6.2.1).
5.1 Preservation.
The ventilator shall be preserved Method IIC in accordance
5.1.1 Level A.
with MIL-P-116
5.1.1.1 Cleaning and drying. Cleaning shall be "C-l", any applicable
process, in accordance with MIL-P-116 and any drying process of MIL-P-116 may
be applied.
Preservative.
No preservative is required.
5.1.1.2
5.1.1.3 Unit packaging. The ventilator shall be cushioned with material
conforming to PPP-C-795 or PPP-C-1842 and shall be placed with an appropriate
amount of desiccant in a sealed bag conforming to MIL-B-117, Types I, II or
III, Classes E, F or G, Styles 1, 2 or 3. The bag shall then be placed into a
PPP-B-636, Type CF, Class Weather Resistant Unit Container.
5.1.2 Commercial. The ventilator shall be preserved and unit packed in
accordance with ASTM D 3951.
Packing shall be level A, B or Commercial (see 6.2.1).
5.2 Packing.
5.2.1 Level A. The ventilator, preserved in accordance with 5.1.1.3 shall
be packed in containers conforming to Table I of MIL-STD-794, as specified for
Level A.
5.2.2 Level B. The ventilator, preserved in accordance with 5.1.1.3 shall
be packed in containers conforming to Table I of MIL-STD-794, as specified for
Level B.
5.2.3 Commercial' The ventilator, preserved and unit packed as specified
above shall be provided packing in accordance with ASTM D 3951.
5.3 Marking. All unit packages and shipping containers shall be marked in
accordance with MIL-STD-129, including bar code markings as specified in
MIL-STD-129 and MIL-STD-1189.
|