| ![]() MIL-V-85830(AS)
First article inspection.
Table II.
I Sample number
123
Test method
Requirement
Inspection
x
x
x
4.8.1
Visual
x
x
x
4.8.2 .1.1
3.9.1.1
Suction leakage
x
x
x
4.8.2. 1.2
3.9.1.2
Pressure leakage
x
x
x
4.8.2.2
3.9.2
Pressure drop
x
x
x
4.8.2.3
3.9.3
Back pressure
x
x
x
4.8.2.4
3.9.4
Insulation Resistance
x
x
x
4.8.2.5.1
3.9.5
Pressure and flow/external power/ground
x
x
x
4.8.2 .5.2
3.9.5
Pressure and flow/battery power/ground
x
x
x
4.8.2 .5.3
3.9.5
Pressure and flow/external power/10000 ft
I
1
I
Table III. Sample size, acceptance criteria, tests
and examination of the ventilator.
Acceptance criteria
Sample size
Section
Inspection
1
For major AQL 1.0%
Normal level II S
4.8.1
Visual
For minor AQL 2.5%
No defects
100%
4.8.2.1
Leakage
AQL 0.657.
Normal level II S
4.8.2.2
pressure drop
acceptance number O
AQL 0.65%.
Normal level II S
4.8.2.3
Back pressure
acceptance number O
AQL 0.65%
Normal level II S
4.8,2.4
Insulation
resistance
AQL 0.65%
Normal level 11 S
4.8.2.5
Pressure/flow
AQL 2.5%
Normal level II S
4.8.1.2
Packaging
L